|
Allen Furst
Beckman Coulter, Inc.
The analytical ultracentrifuge permits observations of the behavior of
macromolecules subjected to a centrifugal field. Such experiments permit determination of
solution molecular weights, association constants, and studies of homogeneity, shape, and
other molecular parameters. The instrument consists of a centrifuge and a rotor with a
windowed sample compartment, so that the distribution of macromolecules along the radial
dimension can be determined at any time by means of an appropriate optical system.
The main advantages of the analytical ultracentrifuge are 1) that it
permits studies of molecules at varying concentrations in many buffers including widely
varying salt and pH conditions, and 2) that the conclusions reached are based directly
upon first principles, and do not require comparisons to standards, which must be assumed
to behave in like manner to the molecules under investigation.
Two basic types of experiment can be performed with the analytical
ultracentrifuge. In a sedimentation equilibrium experiment, a steady state
condition is allowed to develop in which a sample's tendency to sediment in the
centrifugal field is counterbalanced by its tendency to diffuse against the concentration
gradient so established. This is the preferred method for accurate molecular weight
determinations and for studies of homogeneity and molecular associations.
In a sedimentation velocity experiment, the speed with which a
molecule moves toward the bottom (outermost boundary) of the cell is determined. This
yields the sedimentation coefficient, s, which can be related to the molecular
weight and to the frictional coefficient or shape of a particle. In an analogous method,
the flotation velocity (i.e., towards the center of rotation) of materials having
densities less than the solvent is measured. These methods are used for rapid estimates of
the molecular weight and for determining particle size distributions in a sample.
Features of the Optima XL-A
Centrifuge and Rotor
The Optima XL-A ultracentrifuge is readily convertible between analytical and preparative
configurations. The induction drive is equipped with a dynamic damping capability that
reduces rotor precession to near zero. Rotor temperature is monitored by radiometry and
regulated by thermoelectric modules. This system provides rotor temperature stability to
within 0.13°C at equilibrium. The titanium rotor is designed to run double-sector sample
cells, and includes a counterbalance with reference holes for radial calibration.
Absorption Optical System
The optical system is shown in Figure 1. A xenon flashlamp serves as the light source,
providing a usable wavelength range of 190-800 nm. The lamp is fired as the sector of
interest passes over the detector, with the timing regulated by monitoring the passage of
a reference magnet in the bottom of the rotor. The maximum firing rate is 100 flashes per
second, corresponding to once per ten revolutions at 60,000 rpm.

Figure 1. Schematic of the absorbance optical system of
the Optima XL-A. Features of the optical system are described in the text.

A toroidally-curved diffraction grating selects single-wavelength light
for projection onto the sample. The grating is rotated by a precision geartrain to select
the wavelength of interest, and provides a nominal bandpass of 2 nm. A series of absorbing
filters is also provided to block out light of other wavelengths, thus reducing stray
light.
Since the intensity of light from the flash lamp varies somewhat from
pulse to pulse, light from the diffraction grating is normalized by reflecting a small
percentage onto a detector located at the virtual focal point of the monochromator system.
Monochromatic light passes through the sample cell, which is bounded by
two quartz windows. This cell contains both a sample sector and a solvent sector so that
the intensity of light transmitted through the sample can be expressed with reference to
the solvent, as measured by a photomultiplier tube positioned beneath the rotor. A
lens-slit assembly moves as a unit to provide radial scans of these sectors. Multiple
readings can be acquired at each radial distance and averaged to reduce noise. Readings at
several wavelengths can also be taken at each radial distance.
The absorption optical system of the XL-A is based upon the fact that
many macromolecular solutes include chromophores that absorb incident radiation at
particular wavelengths. For solutes obeying Beer's law, the absorption is linearly related
to the molecular concentration. Thus, the radial distribution of the solute of interest,
or C(r), is readily determined from a radial scan of optical density.
Absorption optics offer very high sensitivity. This allows the study of
dilute solutions in which thermodynamic nonideality will be minimal. In typical salt
concentrations (0.1 to 0.2 molar), nonideality can be ignored for all but the most
asymmetric molecules.
A further advantage of absorption optics is the ability to discriminate
between molecules with different chromophores. Thus, the radial distribution of a typical
protein absorbing at 280 nm can be distinguished from that of a potential ligand if the
latter has an absorption peak in the near UV or visible range. This permits
characterization of the stoichiometry and association constant of the binding reaction.
Sedimentation Velocity
Mathematical Theory
Several excellent treatments of this topic are available in the literature (Svedberg and Pedersen, 1940; Schachman, 1959). What follows is a concise overview
(see Figure 2). A particle of mass M in a centrifugal field generated by a spinning
rotor is subjected to a centrifugal force
where omega is the angular velocity in radians per second and r
is the distance in millimeters from the center of rotation. As this particle is moved
through the solvent medium, it displaces solvent molecules and encounters an opposing
force due to buoyancy.
where rho is the solvent density and v-barb the partial
specific volume of the particle (the inverse of particle density). In addition, there will
be frictional resistance to the motion of a particle through the solvent, with the
frictional force equal to
where f is the frictional coefficient, and v the velocity of the
particle. At constant rotor speeds, the velocity of the particle will be constant, and
Substituting and solving for the velocity,
The velocity is generally expressed in terms of the particle
sedimentation coefficient
and expressed in Svedberg units (S; 1 S = 10-13 seconds).
This yields the basic equation applied to sedimentation velocity experiments, namely
Thus, the sedimentation velocity of a particle will increase with its
mass, density, or with the rotational speed, and will decrease with increasing friction
(related to particle asymmetry) or solution density.

Figure 2. Forces experienced by a particle in the
centrifugal field. A particle, shown here in a sector-shaped cell, experiences three
forces during centrifugation. These are the centrifugal force (Fcent), a force
due to buoyancy (Fbuoy), and frictional force (Ffrict). The combined
action of these three forces results in a constant velocity of the particle in the radial
direction.

Experimental Considerations
For sedimentation velocity experiments it is essential that the cells
holding the sample and solvent be sector-shaped, with walls aligned along the radii
of the rotor. This prevents sedimenting particles from colliding with the walls (wall
effects). The ability to resolve boundaries is proportional to omega2rL/theta
where L is the column length, and theta is the width of the section (Svedberg and Pedersen, 1940). Thus, a long,
narrow solution column is generally preferred for highest resolution. Typical solution
columns for sedimentation velocity runs hold 0.45 mL.
Distribution of Concentrations
Sedimentation velocity experiments are normally of a type called boundary sedimentation.
The experiment begins with the sample mixed uniformly throughout the cell, so that a plot
of concentration vs radius is a horizontal line (C(r) = constant). As sedimentation
proceeds, molecules are depleted from the top of the solution column. This results in the
formation of a trailing boundary for the concentration distribution. In an alternative
procedure, termed zonal sedimentation, a special sample cell is used to introduce
the sample to the top of the solution column during the centrifuge run. This results in a
discrete zone of sample molecules migrating through the solvent column.
Figure 3 shows typical data acquired during a boundary sedimentation
experiment. This is simply a plot of the solute concentration as a function of radial
distance, or C(r). Several key features of the data are pointed out in the figure.
A pair of sharp peaks indicate the positions of the menisci. The sample compartment of the
cell is normally filled with slightly less liquid than the solvent compartment, so that
the inner meniscus is that of the solvent compartment and the outer meniscus is that of
the sample compartment. An unusually large distance between these menisci, or one that
shifts during a run, is an indication of leakage from the cell.

Figure 3. Features of boundary sedimentation data. The
figure shows typical data from a boundary sedimentation experiment in the Optima XL-A. The
data represent the absorbance of the fluid in the sample sector compared to the reference
sector of the double-sector cell. Sharp peaks result from the refraction of light away
from the photomultiplier by the menisci in each sector. The sedimentation coefficient is
determined from the movement of the boundary region with time.

The remainder of the data consists of the boundary region in
which the solute concentration increases rapidly to a reasonably constant value in the plateau
region. Most of the information in a sedimentation velocity experiment is taken from
analysis of the boundary. In a simple sedimentation involving one component, the boundary
will be sharp, and the sedimentation coefficient can be derived from the motion of the
boundary midpoint (see below). In more complex analyses involving two or more components,
the boundary will also be divided into two or more rising segments. Assuming each
component has the same extinction coefficient, the relative heights of the boundary
segments return the relative concentration of each component. The radial motion of each
segment can be analyzed independently to determine each sedimentation coefficient. At the
extreme, a very broad boundary is indicative of a heterogeneous sample.
As an alternative representation, the data may be presented as the
derivative of the concentration function, or dC/dr. In this representation, each
boundary segment appears as a discrete peak, the sedimentation coefficient is obtained
from the radial motion of these peaks. The relative concentration of each sample component
is determined from the area under each peak.
One feature of the plateau region is worth noting. Particles at greater
radii will move faster than those closer to the center of rotation, thus pulling away from
the latter. In addition, as the experiment progresses, particles beginning near the
outermost portion of the solution column will be pelleted against the outer wall of the
sample cell, and will be replaced by particles from nearer the center of rotation. These
latter particles enter a progressively increasing volume as they migrate outward through
the sector-shaped cavity, and thus become more dilute. This phenomenon of radial
dilution (Figure 4) accounts for the gradual decrease in optical density in the
plateau (Trautman and Schumaker, 1954).

Figure 4. Radial dilution. These data represent
sedimentation of a 2 S particle at 60,000 rpm during a run of approximately 3 h (10,000
S). As the particles sediment and the boundary moves to the right, the meniscus eventually
becomes depleted (seen at left of figure). In addition, as the particles enter the
increasing volume at the bottom of the sector-shaped cavity (right of figure), the
concentration and the absorbance of the plateau region are seen to decrease.

Data Analysis
Follow the Midpoint
For all but the smallest monodisperse solutes, the sedimentation coefficient may be
obtained by following the rate of motion of the boundary midpoint, rb.
This is most readily calculated from the slope of the equation
where the time, t, is plotted in seconds.
This approach will lead to incorrect results for solutes with s less
than about 2 S. For such solutes, it is preferable to use the second moment method of Goldberg (1953). This requires integration of the
concentration function C(t)rdr from the meniscus to the plateau.
Concentration Dependence
The sedimentation coefficient should be obtained over a range of solute concentrations and
extrapolated to infinite dilution. Highly asymmetric molecules, or molecules forming
associating systems, will show concentration dependence of s (Rowe, 1977).
Molecular volume and extension.
Highly asymmetric molecules tend to occupy a disproportionately large volume due to their
rotational motion in solution. The net effect is to prevent solvent molecules from
approaching them, increasing the apparent viscosity of the solvent and reducing the
sedimentation rate of the asymmetric solute. Thus, the observed sedimentation coefficient
for a highly asymmetric molecule can decrease precipitously with increasing concentration.
Associating systems.
Where macrosolute molecules tend to dimerize or form higher order associating systems, the
sedimentation coefficient may increase with solute concentration. Such associating systems
are best studied by methods of sedimentation equilibrium.
Johnston-Ogston effect for mixtures.
A similar effect is observed in the sedimentation of mixtures (Johnston and Ogston, 1946). At high
concentrations, fast-moving macrosolute molecules must move through a layer of slow-moving
macrosolutes as well as solvent. The slow-moving species increases the apparent viscosity
of the solvent, again leading to a concentration-dependent decrease in the sedimentation
coefficient. This effect is particularly evident for asymmetric molecules.
Speed Dependence
It is also good practice to check for speed dependence in the obtained
values of s. Speed dependence is sometimes observed when sedimenting very large,
asymmetric molecules, or highly polymerized, but dissociable molecules. Some of the causes
are described here. Where speed dependence is observed, it can be avoided simply by
working at the lowest practical rotational speeds.
Speed-dependent aggregation.
The observed sedimentation coefficient for some solutes may increase with increasing rotor
speed. This phenomenon is believed to result from a wake left behind macrosolutes moving
through the solution column, clearing buffer and salt molecules from the medium. This
permits an increased velocity for trailing macrosolute molecules, resulting in the
formation of macromolecular aggregates.
Speed-dependent distortion of large DNAs.
For highly asymmetric molecules, such as DNA, the sedimentation coefficient may appear to
decrease with increasing speed. This is especially true for very large DNA fragments. This
phenomenon is believed to be due to distortion of the molecule at high speeds caused by
solvent friction.
Solvent Effects
Charge.
When charged macrosolutes, such as proteins or nucleic acids, are centrifuged through a
polar solvent, they move more rapidly than the solvent counterions that normally envelope
them. This results in a charge separation and potential difference that slows the
macrosolute molecules, and results in a decrease in the observed sedimentation
coefficient. This effect is generally avoided by using ionic strength in excess of 50 mM.
Viscosity and density.
Very dense or viscous solvents will reduce the observed sedimentation coefficient by
increasing the forces of buoyancy and frictional drag experienced by the macrosolute.
Sedimentation coefficients are therefore conventionally expressed in terms of a standard
solvent, viz., water at 20°C. An observed sedimentation coefficient can be
corrected to the standard, s20,w value with the equation:
where rhoT is the solvent density and rho20,w
that of water at 20°C, etaT/eta20 is the
relative viscosity of water at the temperature T with respect to 20°C, and eta/eta0
is the relative viscosity of the solvent with respect to water.
Applications of Sedimentation Velocity
Analysis of Polydispersity
If a polydisperse solution is composed of particles large enough that there is no
appreciable spreading of the boundary due to diffusion, then the boundary spreading that
is observed will be solely attributable to the spread of s values within the
sample. Under these circumstances, a distribution function g(s) can be defined such
that g(s)ds will equal the weight fraction of macrosolute with sedimentation
coefficients between s and s + ds (Signer and Gross, 1934; van Holde and Weischet,
1978). The function g(s) will depend upon dC/dr, r, t, and the angular
velocity, omega.
Analysis by g(s) can be extended to cover a very broad range of
particle sizes by means of a gravitational sweep experiment, in which the rotor
speed is progressively increased during the run.
Molecular Weight Estimated from s and D
Estimates of the molecular weight of a macrosolute can be determined by sedimentation
velocity experiments. More accurate values are obtained from sedimentation equilibrium
runs. The advantage of the velocity procedure is that it permits simultaneous
determination of the molecular weights for several components in a mixture, providing that
the boundaries formed by these components can be well separated during the course of the
experiment.
The molecular weight, M, can be determined from the following
equation,
also known as the Svedberg equation. Here, R is the gas constant,
T the temperature, v-barb the partial specific volume of the solute, D
is its diffusion coefficient, and rho is the solvent density.
The diffusion coefficient may be obtained through separate analytical
ultracentrifuge experiments in which the spreading of an artificial boundary is observed.
Alternatively, both s and D may be estimated simultaneously by nonlinear
regression (Holladay, 1979).
There are several ways to obtain the partial specific volume. It can be
estimated from the sum of the partial specific volumes of the constituent amino acids (Cohn and Edsall, 1943). Accurate measurements of v-barb
can be made by careful weighings of known solution volumes in a pycnometer. Finally, v-barbrho
can be measured in the analytical ultracentrifuge by determining the quantity M(1 -
v-barbrho) in solvents of different density, such as normal and deuterated
water (Edelstein and Schachman, 1967). An
estimated value of v-barbrho is sometimes used, typically 0.735 mL/g. It should
be noted, however, that a small error in v-barbrho will lead to a considerably
greater error in the determination of s or M.
Determination of Diffusion Coefficient
The diffusion coefficient, D, can be determined from the
spreading of the boundary during a sedimentation velocity run. It is necessary that the
sample consist of a single species or, if multiple species are present, that each boundary
be completely resolved during the run. Details of this analysis are given by Baldwin (1957).
References
- Baldwin, R. L. Boundary spreading in sedimentation velocity
experiments. 5. Measurement of the diffusion coefficient of bovine albumin by Fujita's
equation. Biochem. J. 54, 503-512 (1957)
- Cohn, E. J. and Edsall, E. T. Proteins, Amino Acids, and
Peptides as Ions and Dipolar Ions, pp. 374-377. New York, Reinhold Publ. Co., 1943.
- Edelstein, S. J., Schachman, H. K. The simultaneous
determination of partial specific volumes and molecular weights with microgram quantities.
J. Biol. Chem. 242, 306-311 (1967)
- Goldberg, R. J. Sedimentation in the Ultracentrifuge. J.
Phys. Chem. 57, 194-202 (1953)
- Holladay, L. A. Molecular weights from
approach-to-sedimentation equilibrium data using nonlinear regression analysis. Biophys.
Chem. 10, 183-185 (1979)
- Johnston, J. P., Ogston, A. G. A boundary anomaly found
in the ultracentrifugal sedimentation of mixtures. Trans. Faraday Soc. 42, 789-799
(1946)
- Rowe, A. J. The concentration dependence of transport processes:
a general description applicable to the sedimentation, translational diffusion, and
viscosity coefficients of macromolecular solutes. Biopolymers 16, 2595-2611 (1977)
- Schachman, H. K. Ultracentrifugation in Biochemistry.
New York, Academic Press, 1959.
- Signer, R., Gross, H. Ultrazentrifugale
Polydispersitätsbestimmungen an hochpolymeren Stoffen. Helvet. Chim. Acta 17,
726-735 (1934)
- Svedberg, T., Pedersen, K. O. The Ultracentrifuge.
Oxford, Clarendon Press, 1940.
- Trautman, R., Schumaker, V. Generalization of the
radial dilution square law in ultracentrifugation. J. Chem. Phys. 22, 551-554
(1954)
- van Holde, K. E., Weischet, W. Boundary analysis of sedimentation-velocity experiments
with monodisperse and paucidisperse solutes. Biopolymers 17, 1387-1403 (1978)
|